Elemental Analyzers

Yanaco CHN Corder MT-5, MT-6 (manufactured by Yanaco Apparatus Development Laboratory Co., Ltd.)

Micro Corder JM10, JM11 (manufactured by J-SCIENCE LAB Co., Ltd.)

<Yanaco CHN Corder MT-5 (left) and MT-6 (right)>

<Micro Corder JM10 (left) and JM11 (right)>


《Principal of the analytical method》

The analytical method was based on differential thermal conductivity method. A sample is completely burned under He with O2 gas at a combustion furnace where CuO is filled. The sample is converted to H2O, CO2, NOx gases, and others. NOx gas is reduced to N2 by Cu at a reduction furnace. The other gases except for H2O, CO2, and N2 are removed by absorbers in combustion or reduction furnace. H2O, CO2, and N2 gases are detected by each thermal conductivity detector as counter signals. Finally they are converted to content mass percentages.


Ultra Micro Balances

CAHN CAHN C-32
Mettler Toledo XP2U and XPR2U

CAHN C-32 XP2U


The Service Center of the Elementary Analysis of Organic Compounds, Faculty of Science, Kyushu University